Scans Intermittently abort with EOS Error Code 2252713 is Fixed with 12.0 M5B Software Release for Excite HD and with 15…

Scans Intermittently abort with EOS Error Code 2252713 is Fixed with 12.0 M5B Software Release for Excite HD and with 15.0 M4 Software Release for Signa HDxT systems.

Symptom

Scan Aborts with EOS Error Openspeed system running HFO3 M4. When customer uses the frfse-rpc PSD (intermittently) they will only get half the images reconstructed (only odds). This is because the scan has aborted. Then when they try to download the next series the TPS hangs and the customer has to reset TPS or reboot the system. Operator sees the message "Pulse Sequence Error Occurred". Message log has EOS and other errors logged. See attachments for the review screens and errors logged during these events. MRIhc02494 Openspeed system running HFO3 M4. When customer uses the frfse-rpc PSD (intermittently) they will only get half the images reconstructed (only odds). This is because the scan has aborted. Then when they try to download the next series the TPS hangs and the customer has to reset TPS or reboot the system. Operator sees the message "Pulse Sequence Error Occurred". Message log has EOS and other errors logged. See attachments for the review screens and errors logged during these events. MRIhc02494

Error message

2218537 Scan stopped, Scan Stopped due to error : PSD programming error. Pulse generation was still active when an SSP EOS was received Sat Nov 1 08:42:22 2008 2218537 1 lx-mr scn MsaScanMgr.m 2852 Scan stopped, Scan Stopped due to error : PSD programming error. Pulse generation was still active when an SSP EOS was received..

2252713 PSD programming error. Pulse generation was still active when an SSP EOS was received Wed Sep 5 10:09:30 2007 Error: 2252713 Host: EMR1 Process: NSP File: AGP:seq_cntl.c Line: 671 PSD programming error. Pulse generation was still active when an SSP EOS was received.

Applies to

Models
Signa, Excite
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.