SG20 Fluoro UPS Power does not transfer to Innova during loss of utility power.

Symptom

SG20 Fluoro UPS Power does not transfer to Innova during loss of utility power. Test on PDB is OK. However, when an actual power failure occurs at PDB input, both Innova system and fluoro UPS are switched off. SG20 Fluoro UPS Power does not transfer to Innova during loss of utility power. Test on PDB is OK. However, when an actual power failure occurs at PDB input, both Innova system and fluoro UPS are switched off.

Error message

Errors 410012 410001 410011 and 410018 along with Detector power supply and positioner errors 410012 UPS error. Check UPS error log UPS rectifier main is lost, transition on backup power sytem. 410011 Load on Inverter (OFF) 410001 UPS rectifier main is lost, transition on backup power sytem 410018 UPS power switch failure on switch status Main

Applies to

Models
Innova
Product
Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
Component
SG-20 KVA UPS SG-20 KVA UPS Seelong. SG-20 KVA UPS OPM_SGS_USM_10K_40K_0US_V010.pdf 5380482 UL PDB Phase Control Relay LS1