GE Healthcare · MRExcite225230622246662252357
T/R Driver Fault Due To a Shorted Body Hybrid Splitter.
Symptom
Unable to Scan Head or Body with TR Driver Faults
Error message
2252306 TR Driver Fault Detected Short Circuit Source of Fault is Body TR 2252306 TR Driver Fault Detected - Short Circuit. Source of Fault is Body TR. Value: 123 mA Limit: 80 mA-> 2224666 Auto Prescan failed, Driver Module Error Occurred. Wed Apr 21 20:00:24 2004 Error: 2224666 Host: lx-mr Process: psc File: PscMainControl.cc Line: 1334 Auto Prescan failed, Driver Module error occurred. 2252357 Driver Module error occurred. 2252357 1 lx-mr NSP APS:aps_fastta 135 Driver Module error occurred.
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 5219884-2 NMR Teledyne Hybrid Splitter Kit Upgraded from 46-321312G3 Upgraded from 46-321312G3
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.