How to, install E4500AL - AM x-ray warning/room light control panel and x-ray warning control panels
Symptom
How to, install E4500AL - AM x-ray warning/room light control panel and x-ray warning control panels How to, install E4500AL/am x-ray warning/room light control panel and x-ray warning control panels E4500AL/AM X-RAY WARNING/ROOM LIGHT CONTROL PANEL and X-RAY WARNING CONTROL PANELS Copyright 2002 General Electric Company 1 of 2 Product Data E4500AL/AM 4/17/02 Applications E4500AL, X-ray warning and room lighting control panel provides a low voltage interface between the room warning lights, interior room lighting systems, and the x-ray system. Convenient, pre-wired foot switch operation of the room lights, aids in precise imaging system positioning. E4500AM, X-ray warning light control panel provides a low voltage interface between only the room warning lights and the CT or X-ray system.Interconnection terminals within the imaging system provide the active control signal utilized by this panel. E4500AL panel provides low voltage, low energy control of the room and warning lighting. E4500AM provides only the X-ray warning light on-off control. Low voltage control eliminates the hazard of facility lighting system distribution voltages in the imaging control system cabinets. Servicing safety is improved by the elimination of the 120 or 240 volt lighting circuit from the imaging control system cabinet. Saves time previously used to check, locate and lock out the room warning light and room light circuits. Low voltage control also eliminates the lighting circuit short circuit energy from the imaging control panel. Panel utilizes a dry relay contact in the imaging system to activate x-ray lights via the selfcontained contactor. Installation time, space, and money are saved with the utilization of this UL\cUL listed, CE Labeled, factory manufactured and tested panel. Custom wiring diagram and instruction manual is included for easy coordination of wiring to x-ray or CT Systems. The x-ray Lighting Control Panel provides stock availability of otherwise special order devices, saving time and installation costs. Standard Applications E4500AL - RF and Vascular with room light control and x-ray on warning light. E4500AM - Rad, RF, Vascular, CT with x-ray on warning light. Benefits • Reduces installation time, procurement time and cost, by providing stock availability of this assembled control panel. • Eliminates the sourcing inconveniences and delivery delays often associated with acquiring individual components. • Coordinated and tested to provide the desired functionality. • UL, cUL and CE labeled for compliance to NEC and Global acceptance by inspection authorities. • In stock...available now!
Applies to
- Models
- Innova
- Product
- Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor