Pular para o conteúdo principal
GE Healthcare · MRSignaExcite

Intercom Squeal Caused by Changes to Audio Driver Codes in 15.0 Software.

Symptom

Intercom Noise Producing High Pitch Whining or Squeal Out of the Patient Speaker. IF you see this on an HDxT system, write a Trackwise report. IF you see this on an HDxT system, write a Trackwise report.. The intercom produced a whine at the patient speaker. Replacing the WIM, SCIM, Remote Intercom board (rear pedestal) and associated cables did not fix the problem. We even tried laying out a new set of hardware on the floor outside of the screen room, which bypassed the Pen panel filter. We still had the whining speaker.

Applies to

Models
Signa, Excite
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues