Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnovaAdvantx600465012350125

Innova, Gantry issues, detector lift backs out on its own or May have clicking noise form lift.

Symptom

Innova, Gantry issues, detector lift backs out on its own or May have clicking noise form lift. gantry issues. detector lift backs out on its own or May have clicking noise form lift. Innova, Gantry issues, detector lift backs out on its own or May have clicking noise form lift.t.

Error message

60046 followed by 50123 followed by 50125 60046 POS : SMART HANDLE CONTROL PROBLEM. CHECK IT OR CALL SERVICE. WR POS 0 0 HWStatus=0001 User=0000 0000 4B07 followed by: 50123 POS : TSUI: PERMANENT VALIDATION WR POS 0 0 HWStatus=0002 User=0000 0000 4AC8 followed by: 50125 POS : TSUI: SMB OR SMH ERROR DO POS 0 0 HWStatus=00C7 User=0000 0000 4ACA

Applies to

Models
Innova, Advantx
Product
Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova 2000 Innova 2000 systems Innova 2000 systems
Component
smart handle smart handle smart handle one type of handle is: 5145783-2 FRU Smart handle Zenith_Unity 212000947 09/22/2010 10:28 GMT see vascular suport central site (@vascparts) for a cross refference of handles, boxes, and tssc's 212000947 04/09/2010 18:48 GMT Positioner Gantry Positioner Gantry Positioner Gantry
Area
X-ray Vascular