420V ACGD Breaker Trips Due To a Bad SGA Power Supply. 420 Volt ACGD Circuit Breaker On The PDU Is Tripped OFF.
Symptom
420 Volt ACGD Circuit Breaker On The PDU Is Tripped OFF Important safety issue. Do not have customer reset the 420 Volt ACGD breaker on the PDU cabinet.
Error message
2244587 (breaker tripped) The GP is reporting SGA Power Supply Power Off. Check 420V power breaker on PDU. If breaker is tripped, replace SGA-PS. Check that 420V power line is fully seated between PDU & SGA-PS. Push EMO Reset Switch on PDU. Else, replace SGA-PS. 2224666 Auto Prescan failed, Gradient amplifiers are not ready
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- SecondOpinion ACGD SGA Power Supply 2334862-2 This part requires second opinion approval before ordering. See comments in resolution concept for second opinion requirements before ordering this part.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.