Fails TPS Reset when booting up to Applications or when Resetting TPS.

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2227556 *** Reset/Download TPS was not successful. *** 2227556 sdm sdm_main.c 320 *** Reset/Download TPS was not successful. ***

2249126 SRI Communication Error. A timeout occurred when trying to receive an SRI packet. Mon Apr 21 07:30:58 2008 Error: 2249126 Host: MCMR Process: NSP File: SCP:SriLinkResources Line: 594 SRI Communication Error. A timeout occurred when trying to receive an SRI packet. Opcode = 0xb0

2218661 TPS Communication failed, please try again. See Long Description Error: 2218661 Host: LHMR1 Process: scn File: MsaResearchFlag.m Line: 441 TPS communication failed, please try again.

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2294300-19 MGD SCPII with CAN PMC and I2C firmware fix 2294300–19 SCP2 BOARD
Area
MR