Pular para o conteúdo principal
GE Healthcare · MRExcite226486222460272264861226506222238172264859

CWA ID140741 Cannot scan with DCERD errors.

Symptom

Cannot scan with DCERD errors

Error message

2264862 The fiber link between the DCERD and MGD chassis is not in a ready state 2264862 Host: hd01 Process: diagCoord File: Irf.cpp Line: 290 Task ID: 0x7effdf0 Time: 1835 Object: APS Exception Class: Unknown Function: RF: RF Transmit RRF FIBER CABLE LINK NOT READY The fiber link between the DCERD and MGD chassis is not in a ready state after hardware initialization. This will prevent the MGD from controlling the DCERD chassis. Try TPS reset. Run DCERD diagnostics.

2246027 The diagnostic platform was accessed with an invalid parameter

2264861This error message isn't supported for DCERD. If this message displayed, it may be software problem Error: 2264861 Host: hd01 Process: diagCoord File: Irf.cpp Line: 905 Task ID: 0x7effdf0 Time: 2447 Object: APS Exception Class: Unknown Function: RF: RF Transmit This error message isn't supported for DCERD. If this message displayed, it may be software problem.-

2265062 DCERD Fiber Optic Link Error on the Master DCERD Digital board. Rx Signal Detect signal is in an Invalid State. Extended Error: Check the Gigalink fiber optic connection between the IRF I/O and DCERD Digital Boards. If fiber cables are connected and the system indicates that 

Applies to

Models
Excite
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 14.0 For Hino Excite HDe 14.0 systems only.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues