BrightSpeed Select or Delight Select
Symptom
Scanner Hardware Reset Failed. See LD Scanner Hardware Reset Failed Hardware Failed to reset message appeared on monitor AFTER boot-up WITH Error:178568 Failed to establish socket connection to STC in 10 seconds and errors 244553 and 244554 and power up self test fails A, B, and 9 on STC CPU board The following errors were in the gesyslog Thu Jan 13 04:18:32 2005 Error: 178568 Suite: RP25 Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to STC in 10 seconds Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset. Thu Jan 13 04:18:32 2005 Error: 244554 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 442 The reset failure of the OBC may be due to the reset failure of the STC. Also saw failures A, B, and 9 on the power-up test on the STC CPU .
Error message
Error: 260101012 The Rotating Controller Interface Bus (RCIB) fault line has been asserted. See LD for more details Mon Apr 10 07:20:41 2006 Error: 260101012 Host: orp Process: orp subsys Mon Apr 10 07:20:40 2006 Host : On-Board Rotating Processor Ermes # : 260101012 Exception Class : Abort Severity : Primary File : rcib fault state.c Line# : 139 Function : No System Function Reported Scan Type : Axial Scan: 50060/6/11 The Rotating Controller Interface Bus (RCIB) fault line has been asserted.None of the sub-systems have reported a problem. Check nodes or fault line for failure. Error: 260001514 RCIB Bus: Fault Line Interrupt Occurred. Timeout Waiting for Host Response to Fault Line Verification. See long description for full error Wed Apr 27 01:00:51 2005 Error: 260001514 Host: orp Process: orp_subsys File: notiftsk.c Line: 1368 Running: System Reset Wed Apr 27 01:00:51 2005 Host : Das Ermes # : 260001514 Exception Class : Abort Severity : Secondary File : notiftsk.c Line# : 1368 Function : No System Function Reported Scan Type : None/Unknown Scan: 50004/1/0 RCIB Bus: Fault Line Interrupt Occurred. Timeout Waiting for Host Response to Fault Line Verification. (Timeout:500 msecs).
Applies to
- Models
- BrightSpeed, VCT
- Product
- BrightSpeed Select, Delight Select or MERI Select Elite-16, Edge-8, Excel-4 HP_S_G16_G_HLT HP_S_G8_G_HLT HP_S_G4_G_HLT BrightSpeed Select or Delight Select (Hivalue Lite) 4/8/16 Slice Scanner BrightSpeed Delight Select Systems started to ship in first Quarter or 2009 with AIO (All in One) Console. AIO console uses HP8400 and NO DARC or IG's. Recon time is 6 FPS. If customer orders Fluoro option system requires one IG. The 16 FPS Upgrade in Not Available for Select systems. BrightSpeed Meri Select Systems ( Pancake Das ) started to ship in 3rd quarter of 2010, 4/8/16 Slice Scanner HiValue 16 Lite −−> BrightSpeed Elite Select HiValue 8 Lite−−> BrightSpeed Edge Select HiValue 4 Lite−−> BrightSpeed Excel Select BrightSpeed Select Scope: - 3.5MHu Solarix tube - JEDI 60 dc Lite - Modified NP Table Key Features: - 16x0.625mm(16x1.25) for Lite 16 - 8x1.25/8x2.5mm for Lite 4/8 - Rotation Speed: 0.8 s - Gantry: 2040mm width, 70cm Bore - Table: 1337@ISO-140mm scannable range - Recon: 3fps (w/o IG @ 2G Memory(4/8), 4G Memory(16)), optional: 6fps(w/ 1IG ) - Power: 42KW JEDI power See the attached Service Intro Planner for details Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required