Pular para o conteúdo principal
GE Healthcare · MRSignaExciteDiscoveryOptima2223817

CWA ID140558. TPS not responding during applications boot with Error 2223817.

Symptom

TPS not responding during applications boot with Error 2223817 See long description and attached file. NO errors in mgd_term windows. All three processors finish boot with applications startup complete message. /usr/g/service/log/scn.out file showed MGD chassis failed boot. See attached file.

Error message

2223817 TPS is Not Responding. Please Reset TPS Wed Apr 28 16:58:01 2004 Error: 2223817 Host: GEMS Process: NSP File: TardisReset.cc Line: 398 TPS is not responding. Please reset TPS.

Applies to

Models
Signa, Excite, Discovery, Optima
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution