GE Healthcare · MRExcite225205511217243322224360

Defective IT Fiber Optic cable between the Host computer and the System Cabinet.

Symptom

Some or All Images in the Last Scan Did Not Reconstruct Some or All Images in the Last Scan Did Not Reconstruct.

Error message

2252055 Diagnostic: IT Host->MGD Hammer. IT-Host card driver initialization error. 2252055 lx-mr diagCoord bit3_diags.c 469 Diagnostic: IT Host->MGD Hammer Iteration: 1 of 1 Action: 1 Task ID: 0x0 Time: 1121724332 Object: Host Diag Executive IT-Host card driver initialization error. This error usually means that the IT-Cable is disconnected or that the cable is damaged.->

2224360 Unable to communicate to the Tardis chassis after multiple attempts. Image 7208/1/2 was not installed in the database. Fri Oct 8 07:53:33 2004 Error: 2224360 Host: mr02 Process: TIR re_pixel_header.c Line: 320 Host: mr02 Process: TIR File: store_pixel_header.c Line: 320 Unable to communicate to the Tardis chassis after multiple attempts. Image 7208/1/2 was not installed in the database. Check the Bit-3 connection to the Tardis chassis if further images are lost.

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2318811 IT-Link Cable Run # 1044 OW1 A15 BIT3 TO MR2 A11 J22
Area
MR