GE Healthcare · MRExciteSigna2224864
Unexpected Table Motion Errors Due To Table Out Of Adjustment.
Symptom
Frequent error messages regarding unexpected table motion. There are also a high number of 2248597 (pause button pressed). This causes the scan to stop and the customer needs to press advance to scan.
Error message
2224864 Table motion was stopped because it was moving faster than expected. Tue Jun 2 10:59:30 2009 Error: 2224864Host: IMCMROP02 Process: NSP File: SCP:SRI Line: 99Exception Class: Unknown Severity: UnknownPSD: Gradient Echo Coil: 8CARDIAC Scan: 3806/13Function: Patient Handling: Longitudinal Motion. Table motion was stopped because it was moving faster than expected. This can occur if the table was moved by hand. Limit: 150 mm/sec, Last Velocity Commanded: -100 mm/secMeasured Cradle Velocity: -169 mm/sec->
Applies to
- Models
- Excite, Signa
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 M5 Excite HDx HDi Platforms only.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.