Unexpected Table Motion Errors Due To Table Out Of Adjustment.

Symptom

Frequent error messages regarding unexpected table motion. There are also a high number of 2248597 (pause button pressed). This causes the scan to stop and the customer needs to press advance to scan.

Error message

2224864 Table motion was stopped because it was moving faster than expected. Tue Jun 2 10:59:30 2009 Error: 2224864Host: IMCMROP02 Process: NSP File: SCP:SRI Line: 99Exception Class: Unknown Severity: UnknownPSD: Gradient Echo Coil: 8CARDIAC Scan: 3806/13Function: Patient Handling: Longitudinal Motion. Table motion was stopped because it was moving faster than expected. This can occur if the table was moved by hand. Limit: 150 mm/sec, Last Velocity Commanded: -100 mm/secMeasured Cradle Velocity: -169 mm/sec->

Applies to

Models
Excite, Signa
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 M5 Excite HDx HDi Platforms only.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required