GE Healthcare · MRExcite2244871

CWA ID140116. Will Not Scan with Gradient Errors.

Symptom

Will Not Scan with Gradient Errors.

Error message

2244871 The GP reported multiple Gradient Amplifier % Axis OverVoltage Faults. See Long Description Extended Error : If an Overvoltage fault is not also reported on the SGA Power Supply, replace the Gradient Amplifier. Else replace the SGA Power Supply. TPS reset is required to recover from this fault.

Applies to

Models
Excite
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 M5 Excite HDx HDi Platforms only.
Area
MR