Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova1050252

The Table Rotaion Lock is not working

Symptom

Omega, Unexpected table rotation or rotational Lock is not working. See FMI 12137 Omega IV and Omega V tables may experience an unexpected table rotation during patient transfer due to rotational table brakes slippage. FMI 12137 Dated 1/2010 Lost of voltage on lock will kept the on. Voltage is to remove the brake or hold of the lock. Omega IV and Omega V tables may experience an unexpected table rotation during patient transfer due to rotational table brakes slippage. Lost of voltage on lock will kept the on. Voltage is to remove the brake or hold of the lock. Table top rotation lock not holding. FMI 12137 Dated 1/2010, for all Omega 4 &5 DOC, DOC0683423, Rev:2 PURPOSE / HISTORY This FMI is generated to fix Complaint 13238255 (Master). Omega IV and Omega V tables may experience an unexpected table rotation during patient transfer due to rotational table brakes slippage. This event is most likely to occur when patient is transferring himself from the bed to the table, leading to a potential patient fall. Products labelling specifies that when loading and unloading the patient onto the table, table top should be maintained in order to ensure it does not move. The omega table rotation brakes need to comply to 2 contradictory specifications: 1) Brakes should be strong enough to hold table and avoid unexpected rotation when locked. 2) Brakes should be weak enough to slip allowing to disengage patient if a CPR required while power is off. Investigation showed that table rotation torque specs should be increased from 26.5 to 38.7 daN.m.

Error message

error Code = 1050252, error Text = POSITIONER LOCKED, debug Text = - lock button pressed on SH/SB, recovery Class = WR, major Function /= GantryAxis System impact: Gantry/table lock-up. Displayed message : "Gantry/Table commands disabled. Use Smart Box/Handle to enable.".

Applies to

Models
Innova
Product
X Innova 4100 X Do not use---------4100 systems 4100 systems
Software
$ Unity M4 SP3.2 DL_Unity_4.18.9, FMI 12105 Unity M4 SP3.2 DL_Unity_4.18.9 FMI 12105 Unity M4 SP3.2 DL_Unity_4.18.9, FMI 12105 healthpage tag for this software is UNITY_M4_FMI_4.21 212009270 03/26/2010 13:31 GMT
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare X-ray Vascular issues