CWA ID139457. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
Symptom
Computer will not boot up to applications level, with the message on the screen that says Please do not interact with the system until this message disappears. Computer will not boot up to applications level, with the message on the screen that says -- > Please do not interact with the system until this message disappears. If the system does not come up in 15 minutes, please call your service representative.
Error message
200002352 A HOST_SHUTDOWN was initiated by the death of RTIA Fri Apr 23 06:28:51 2004 Error: 200002 352 Host: twin-mr1 Process: host Monitor File: ssProcStop.c Line: 810 A HOST_SHUTDOWN was initiated by the death of RTIA.
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.