Unable to Scan with HD Lower leg Array Right Coil or Boot due to Error Code 2252303 Multicoil Bias Fault Short Circuit d…

Unable to Scan with HD Lower leg Array Right Coil or Boot due to Error Code 2252303 Multicoil Bias Fault Short Circuit due to Coil's Preamp Box.

Symptom

Unable to Scan with HD Lower leg Array Right Coil or Boot with Error Code 2252303 Multicoil Bias Fault Short Circuit. See Long Description Unable to Scan with HD Lower leg Array right-side coil or boot due to Error Code 2252303 Multicoil Bias Fault - Short Circuit. The errors shown below are logged. The left-side boot works fine and the problem doesn't happen with any other coil. Wed Feb 11 15:16:41 2009 2252293 1 mr1 NSP SCP:DCB 2969 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: Spin Echo Coil: HD Lower Leg R Scan: 50732/1 The Driver Module Has Detected a Fault. Details of the Fault can be found in other error messages.-> Wed Feb 11 15:16:41 2009 2252303 1 mr1 NSP SCP:DCB 1024 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: Spin Echo Coil: HD Lower Leg R Scan: 50732/1 Multicoil Bias Fault - Short Circuit. Try re-seating all coil connectors. MCD1 Channel 9 of 16, counting from 1 Value: -459 mA, Limit: -201 mA Mask of failed channels: 0x100 MCD1 1-8 : A or C MCD1 9-16 : B MCD2 1-8 : B or Legacy MC MCD2 9-16 : C-> Wed Feb 11 15:16:42 2009 2252357 1 mr1 NSP APS:aps_cf 243 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: Spin Echo Coil: HD Lower Leg R Scan: 50732/1 Driver Module error occurred. Wed Feb 11 15:16:42 2009 2224666 1 mr1 psc PscMainControl.cc 1358 Exception Class: Unknown Function: Prescan: Host Autoprescan Auto Prescan failed, possible data acquisition error. Try again or use Manual Prescan.

Error message

2252357 Driver Module error occurred. 2252357 1 lx-mr NSP APS:aps_fastta 135 Driver Module error occurred.

2224666 Auto Prescan failed, possible data acquisition error. Try again or use Manual Prescan. Wed Dec 1 09:09:30 2004 Error: 2224666 Host: lx-mr Process: psc File: PscMainControl.cc Line: 1322 Auto Prescan failed, possible data acquisition error. Try again or use Manual Prescan.

2252303 Multicoil Bias Fault Detected, Short Circuit. Channel X of Y See Long Description Error: 2252303 Host: lx-mr Process: NSP File: SCP:DCB Line: 1030 Multicoil Bias Fault Detected - Short Circuit. Try re-seating all coil connectors. MCD1 Channel 3 of 16, counting from 1 Value: -310 mA Limit: -301 mA Mask of failed channels: 0x3c Ch 1-8 : Legacy MC Conn Ch 17-24 : Conn A Ch 25-32 : Conn B-> Extended Error: This fault was detected by the Driver Module. A short circuit fault occurs when the actual value is more negative than the pre-defined limit. This could be caused by a shorted diode in the coil or a short elsewhere in the system. Check coil, system cables, connectors and configurations. The mask of all failed channels indicates additional channels with the same fault - convert the hex number t

Applies to

Models
Excite
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 M5 Excite HDx HDi Platforms only.
Component
2414069 FRU PREAMP BOX W/O CABLE 3RD STATION 2414069 FRU PREAMP BOX W/O CABLE 3RD STATION
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.