Using the HNS coil Scans Stop Intermittently With Error Code 2248008 "RFM EOS semaphore timeout error" due to IRF2 board…

Using the HNS coil Scans Stop Intermittently With Error Code 2248008 "RFM EOS semaphore timeout error" due to IRF2 board.

Symptom

Scans Stop Intermittently With Error Code 2248008 "RFM EOS semaphore timeout error" See Long Description Scans stop in the middle of the scan with the following error logged. This happens about once per day. (PQR 13096684) 2248008 EXCITE15T NSP APS:DRF 250 RFM EOS semaphore timeout error. IRF did not send enough data to DRF. Please change prescription or perform a TPS Reset. DRF ID: 1 FE ID: 1 segCount: 12 frameCount: 14

Error message

2248008 RFM EOS semaphore timeout error. IRF did not send enough data to DRF.

Applies to

Models
Excite
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 M5 Excite HDx HDi Platforms only.
Component
2358122-2 IRF2 Board
Area
MR