Intermittent CAN link errors due to Driver Module.
Symptom
Intermittent MDS Link Failures while Scanning. Getting MDS link failures while scanning. Reports error 2241608 GIP VCC signal is out of range. Note that the value was mere millivolts out of spec, but enough to trip out scans. I measured the voltage at 4.81 volts, nominally low, but barely in spec. Spec is: Low: 4.70V High: 5.250V.
Error message
2252245 Network Management Recovery detected on the CAN network. The node reset and recovered. Recovered Node is Master Gradient. 2252245 IMIMDLMR2 NSPSCP:CanLinkHWO 812Exception Class: Unknown Severity: UnknownFunction: Peripherals: CommunicationsPSD: Unknown Coil: Unknown Scan: -1/-1Network Management Recovery detected on the CAN network.The node reset and recovered.Recovered Node is Master Gradient.
Applies to
- Models
- Excite
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 M5 Excite HDx HDi Platforms only.
- Component
- SecondOpinion Excite HD & HDx Driver Module 5110612-2 See comments in resolution concept for second opinion requirements before ordering this part
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.