Pular para o conteúdo principal
GE Healthcare · MRSignaExciteDiscoveryOptima222388322238542223859

CWA ID138789. Can Not Boot Application Software With VRE Errors.

Symptom

Can Not Boot Application Software With VRE Errors See Long Description 2249418 Host: LJP MR 01 Process: diagCoord File: mgdArp.c Line: 154 Task ID: 0x7effdf0 Time: 1689 Object: APS Exception Class: Unknown Function: Software/Firmware/Config: N/A Network problem detected. Check MGD subnet network configuration. arpResolve failed for IP 10.0.0.102, numTries 2, delay 100 ticks. status = -1. Or, 2223859 Host: ghmr002 Process: NSP File: TardisReset.cc Line: 469 Exception Class: Unknown Function: Recon: Data Acquisition The TPS Reset failed because of a VRE reset failure. Or; 2223860 VRE reset failed on ICN icn1 with MAC address XX:XX:XX:XX:XX:XX Or, 2223885 Host: hde-mr Process: logermes File: vreBmcUpCheck.pl Line: 251 Exception Class: Shutdown Function: Host: NSP System startup failed due to failure to communicate with the VRE!-> Extended Error: The system startup failure is caused by the failure to communicate with all configured ICNs. Please check that power is applied to all ICNs. Please check the Ethernet connections to all ICNs. You may also run all Ethernet and IPMI tests under VRE Communication Diagnostics to verify connectivity. As a last resort, VRE Configuration may be required to recover.

Error message

2223883 VRE reset failed due to failure to communicate with all ICNs! Check Ethernet and IPMI connectivity to all ICNs See long description Mon Jan 26 08:47:14 2009 Error: 2223883 Host: OVMRI1 Process: NSP File: VREReset.cc Line: 937 Exception Class: Unknown Function: Recon: Data Acquisition VRE reset failed due to failure to communicate with all ICNs! Check Ethernet and IPMI connectivity to all ICNs.-> Extended Error: The TPS reset failure is caused by the failure to communicate with all configured ICNs. Please check the Ethernet connections to all ICNs. You may also run all Ethernet and IPMI tests under VRE Communication Diagnostics to verify connectivity. As a last resort, VRE Configuration may be required to recover.

2223854 Extended TPS Reset in progress (up to 8 minutes). Please be patient.-> See Long Description 2223854 ERROR: EM_mrNSP_VRE_RESET_PROLONGED Extended TPS Reset in progress (up to 8 minutes). Please be patient. Extended Error: One or more ICNs of the VRE might have been non-responsive during a TPS Reset. We therefore reboot all ICNs in an attempt to recover. The operator will experience longer TPS Reset times because of this.

2223859 The TPS Reset Failed becaus

Applies to

Models
Signa, Excite, Discovery, Optima
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues