GE Healthcare · MR
OrthoMRI The PACS (remote) computer disk was full
Symptom
OrthoMRI Unable to send images
Error message
OrthoMRI Out of resources error message See Long Description Notice: Always use the most current revision of service documentation! Out of resources error. This is in reference to the amount of free space on the hard drive of the host computer or the remote computer. You can try to archive then delete images from the system database just to increase the free space in the hard disk. NOTE: There is no way to check the free space on an OrthoMRI system. Also, on the PACS (remote) computer you can check the free space in the hard disk, after this operation you can try again to send the images.
Applies to
- Product
- OrthoMRI EscanNE. OrthoMRI EscanNE.
- Software
- OrthoMRI S/W Release 9.1B. OrthoMRI S/W Release 9.1B.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- OrthoMRI Network cable is unplugged
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- CWA ID100060 Unable To Film.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.