GE Healthcare · MRExciteSigna222385922494182223854
Unable to Reset TPS Due to Defective Ethernet Cable to ICN.
Symptom
Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis
Error message
2223859 The TPS Reset Failed because of a VRE Reset Failure See long description. 2223859 Host: ghmr002 Process: NSP File: TardisReset.cc Line: 469 Exception Class: Unknown Function: Recon: Data Acquisition The TPS Reset failed because of a VRE reset failure. Extended Error: This indicates that a failure to reset the VRE properly caused the TPS Reset to fail. Please refer to the VRE Troubleshooting Guide for further troubleshooting this issue, which can be found in the MR Service Documentation. 2249418 Network problem detected. Check MGD subnet network configuration. See Long Description Wed Apr 14 09:27:10 2004 2249418 2 CARD diagCoord mgdArp.c 157 Task ID: 0xf99df8 Time: 981 Object: SCP Network problem detected. Check MGD subnet network configuration. etherAddrResolve failed for network interface motfec0, IP 10.47.1.2, numTries 5, delay 100 ticks. status = -1. 2223854 Extended TPS Reset in progress (up to 8 minutes). Please be patient.-> See Long Description 2223854 ERROR: EM_mrNSP_VRE_RESET_PROLONGED Extended TPS Reset in progress (up to 8 minutes). Please be patient. Extended Error: One or more ICNs of the VRE might have been non-responsive during a TPS Reset. We therefore re
Applies to
- Models
- Excite, Signa
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 M5 Excite HDx HDi Platforms only.
- Component
- 5160775 FRU ICN Ethernet Cable Collector This kit contains: 3 cables between ICN and Ethernet Switch, 1 J63 to Ethernet Switch cable, 1 Ethernet cable 100 ft (30.5m) in length between Host and J63 on System Cabinet, and 1 cable label book. HDx system.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.