GE Healthcare · MRExciteSigna22486792248683
MDS Configuration failed and RF Cabinet MDS peripheral response timeout due to defective SSM.
Symptom
TPS or MGD Reset Fails With MDS Link Errors
Error message
2248679 RF Cabinet MDS peripheral response timeout Tue Jun 6 06:20:12 2006 Error: 2248679 Host: lx-mr Process: NSP File: SCP:MdsAutoConfig Line: 1830 RF Cabinet MDS peripheral response timeout. Unknown link failure internal to the RF Cabinet. Unknown link: 0x00 Please power cycle the RF Subsys 2248683 MDS Configuration Failed. MDS link errors caused the configuration to fail. Tue May 25 10:38:44 2004 2248683 MROP NSPSCP:MdsAutoConfig 338 MDS Configuration Failed.MDS link errors caused the configuration to fail. Please fix and perform a TPS Reset.
Applies to
- Models
- Excite, Signa
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- SecondOpinion System Support Module 2160200-43 See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.