GE Healthcare · MRExcite2248650

Patient comfort sensor trip Due to Gradient Cooling Problem

Symptom

Intermittent Warnings of high Patient Comfort Temperature Readings Intermittent Warnings of high Patient Comfort Temperature Readings.

Error message

2248650 The patient comfort level is warmer than normal. The patient may be uncomfortable. Further increases in the temperature will inhibit scanning. 2248650 b4a NSP SCP:BoreTempSriHWO 987 The patient comfort level is warmer than normal. The patient may be uncomfortable. Further increases in the temperature will inhibit scanning. Verify: 1. Patient fan is ON. 2. Room temperature is normal. 3. Air flow through the bore is unobstructed.

Applies to

Models
Excite
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 14.0 For Hino Excite HDe 14.0 systems only.
Component
2222564-30 50 and 60 HZ chiller by Lytron. See Long Description 2222564-30 50 and 60 HZ chiller by Lytron Model Number RC045J03BG0C011 Specified by 2222564PSP. Lower level FRU's should be utilized before attempting to replace the entire chiller. Gradient Chiller Motor - 2222564-25 Pump Assembly for Lytron Chiller - 2222564-29
Area
MR

Full cause & step-by-step resolution