Pular para o conteúdo principal
GE Healthcare · MRExcite

Cradle Will Not Advance With Patient Loaded due to Cradle Release Block is Extended.

Symptom

Cradle Will Not Advance with Cradle Obstructed. Moves Okay Without Patient With a patient on the table, not a heavy one, the table will not move in and out, and the system returns a message "Cradle obstructed". There is no unusual noise heard. But with no patient on the table, it moves in and out fine. The customer tried cycling RF/Pen cabinet power but this was no help.

Error message

The Cradle is Obstructed This is a popup error message that the operator sees when the cradle is commanded to move and the system does not sense the movement. (The system counts encoder pulses per unit of time and will fail if none or too few pulses are seen). The error message is not logged in the message log.

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues