GE Healthcare · MRExcite2252305
Head TR Driver faults Due to Loose head heliax connection at the penetration panel.
Symptom
Cannot scan in Head Transmit Mode with Head TR Driver Faults. Body Transmit works OK.
Error message
2252305 TR Driver Fault Detected Open Circuit. Source of Fault is Head TR 2252305 lx-mr NSP SCP:DCB 1240 TR Driver Fault Detected - Open Circuit. Source of Fault is Head TR. Value: 0 mA Limit: 900 mA->
Applies to
- Models
- Excite
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 M5 Excite HDx HDi Platforms only.
- Component
- 5131924 Times Microwave Heliax Cable Repair Kit See Long Description The 5131924 Times Microwave Heliax Cable Repair Kit Contains the following: 2335190 LMR 900 Stripping Tool 2352193 LMR 600 Stripping Tool 2384858 LMR 1200 Stripping Tool 5111565 RF Cable Cutting Tool Rigid # BK 125S
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.