Power Monitor Faults Due to Patient weight was entered incorrectly.

Symptom

Auto Prescan failed, RF Amplifier is not ready Getting another message that RF Amplifier is powering but message never goes away.

Error message

2254782 The Universal Power Monitor 2 has detected Peak Power Exceeded on the Narrow Band Head Amplifier. 2254782 1 df15hdmr NSP SCP:UPM 657 Exception Class: Unknown Severity: Unknown Function: RF: RF Power Monitor PSD: Spin Echo Coil: HEAD Scan: 7975/1 The Universal Power Monitor 2 has detected Peak Power Exceeded on the Narrow Ban d Head Amplifier. Check patient weight entry. Peak value detected = 657 W. Max Peak Limit = 626 W-

2254782 The Universal Power Monitor 1 has detected Peak Power Exceeded on the Narrow Band Head Amplifier. 2254782 NSP SCP:UPM 657 The Universal Power Monitor 2 has detected Peak Power Exceeded on the Narrow Band Head Amplifier. Peak value detected = 429 W. Max Peak Limit = 426 W- Possible cause(s): ASC RF Detector, ASC RF Processor, RF Amp, Software .

Applies to

Models
Excite
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 For both Excite HDx and 3T Excite HDx systems.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.