Independent Vector Gating unavailable in scan application due to PAC Type is set incorrectly.
Symptom
Independent Vector Gating unavailable in scan application. "Independent Vector Gating" is the new Cardiac Gating option starting with Excite HD 12.0 software. But the technologist is unable to select it in the Scan Rx. It is unavailable.
Error message
2226158 The system detected a PAC II connected. Please check the PAC version connected to the system. Vector Gating will ONLY operate reliably with PAC III version.
Applies to
- Models
- Signa, Excite
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.