The Scan Stops Because the Application Did Not Receive a Trigger on Time due to Defective PAC.
Symptom
ECG Cardiac Gating and PPG Pulse Gating lose their Waveform as Soon as Scan Starts Customer also had a third party Invivo monitor for cardiac monitoring, the waveform on the Invivo monitor did not drop out during the scan
Error message
2226158 The system detected a PAC II connected. Please check the PAC version connected to the system. Vector Gating will ONLY operate reliably with PAC III version.
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 5112668-2 PAC–2B MODULE PAC–2B MODULE 5112668-2 PAC–2B MODULE 5112668-2
- Area
- MR