Pular para o conteúdo principal
GE Healthcare · MRExcite110291103311024

LX systems view the Dicom images from the workstation as an Ultrasound image, which is not a compatible image type for L…

LX systems view the Dicom images from the workstation as an Ultrasound image, which is not a compatible image type for LX MR.No Workaround.

Symptom

Unable To Transfer (Pull) Images To Scanner Customer can query an ALL image store system from the LX console, but transfers (get) fail.

Error message

11029 MFM Warning: Error:element is not UI UID string Mon Nov 5 11:38:58 2001 Error: 11029 Host: MRS1 Process: dcserver File: dcsformat.c Line: 1715 MFM Warning: Error:element is not UI (UID string) (-102) Group:0x8 Element:0x18

11033 Unknown modality type US Mon Nov 5 11:38:32 2001 Error: 11033 Host: MRS1 Process: dcserver File: dcsformat.c Line: 1671 Unknown modality type US

11024 Failure to parse FIND response: Continuing! Mon Nov 5 11:38:32 2001 Error: 11024 Host: MRS1 Process: dcserver File: dcslist.c Line: 511 Failure to parse FIND response: Continuing!

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Area
MR

Full cause & step-by-step resolution