GE Healthcare · MRExcite2254782
UPM Body functional checks fail with peak power faults Due to RF Body Out of Calibration.
Symptom
UPM Body Functional Checks Fail with Peak Power Faults UPM Body functional checks fail with peak power faults for all three failure modes which then fails functional check.
Error message
2254782 The Universal Power Monitor 1 has detected Peak Power Exceeded on the Narrow Band Body Amplifier. Sun Nov 20 12:49:19 2005 Error: 2254782 Host: HDMR Process: NSP File: SCP:UPM Line: 657 The Universal Power Monitor 1 has detected Peak Power Exceeded on the Narrow Band Body Amplifier. Peak value detected = 2236
Applies to
- Models
- Excite
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 M5 Excite HDx HDi Platforms only.
- Area
- MR
Full cause & step-by-step resolution