Gantry moving slow due to DETECTOR collision sensor errors.
Symptom
Gantry moving slow due to DETECTOR collision sensor errors. Typically indicated by HW codes 0004 for the detector faceplate or 0008 for the detector bladder. Gantry moving slow due to detector collision errors. This issue is caused by a collision detected on the Detector side of the gantry. Typically indicated by HW codes 0004 for the detector faceplate or 0008 for the detector bladder. System impact: Message displayed on monitors and BIP on TSSC every 30 seconds. Full gantry motion. Displayed message : "Collision detection failure, continue with care; Otherwise call Service" The Hardware code tells whether the problem is located on the Detector or on the Tube side: HW=4 for the Detector Faceplate HW=8 is the Detector Bumper and HW=10 for the Tube . Check the anti collision sensors on the Detector. Touch the collision sensor margin on one side to re-balance the collision sensors. Touch the collision sensor margin on both sides to clear the collision error. If the problem persists (i.e., the error is not cleared), replace the corresponding ANTI-COLLISION COVER
Error message
error Code = 1050306, error Text = TUBE BUMP SENSOR PROBLEM, debug Text = - TUBE COLLISION DETECTION PROBLEM during MP_DiagnosticBumpers (T:0x4AB5 H:0x10 U:0x2), recovery Class = WR, major Function /= GantryAxis System impact: Full gantry motion. Aborts Bolus and 3D acquisition. Displayed message : "Tube collision detection error. Continue with care or call Service." Or since Unity M4 : "Tube collision sensor error. Press Tube front plate to re-init." System impact: Full gantry motion. Aborts Bolus and 3D acquisition. Displayed message : "Tube collision detection error. Continue with care or call Service." Or since Unity 3D M4 : "Tube collision sensor error. Press Tube front plate to re-init."
Applies to
- Models
- Innova
- Product
- Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
- Component
- Positioner Gantry Positioner Gantry Positioner Gantry
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.