GE Healthcare · X-ray VascularInnova13001071300112
Xray not available; Reset system to recover, Call Service is displayed in the Image monitor.
Symptom
Xray not available; Reset system to recover, Call Service is displayed in the Image monitor.
Error message
INX1X1: 1300107 Lat: Boot critical conditioner failure Failed to boot the conditioner. Lateral Plane 1300107 Lat: Boot critical conditioner failure Failed to boot the conditioner. Error Text: 1300107 IH XRImDet Lat: Boot critical conditioner failure Debug: [DemSupervisor::_bootConditioner] Failed to boot the conditioner This error will always be preceded by warning 1300112 DO XRImDet Lat: Boot critical conditioner warning. Root Cause: This error is due to time out on Communication to the lateral Detector Conditioner or bad status returned by Conditioner. 1300107 Lat: Boot critical conditioner failure Failed to boot the conditioner. Error Text: 1300107 IH XRImDet Lat: Boot critical conditioner failure Debug: [DemSupervisor::_bootConditioner] Failed to boot the conditioner This error will always be preceded by warning 1300112 DO XRImDet Lat: Boot critical conditioner warning. Root Cause: This error is due to time out on Communication to the lateral Detector Conditioner or bad status returned by Conditioner.
Applies to
- Models
- Innova
- Product
- Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
- Software
- $ M4 SPAD BIP_4.18.5, DL_SPAD_4.15.1 FMI 12108 $New, M4 SPAD BIP_4.18.5, DL_SPAD_4.15.1 FMI 12108 or $ new, Innova Biplane M4 BIP_4.18.5, DL_4.15.1 FMI 12108 $New, M4 SPAD BIP_4.18.5, DL_SPAD_4.15.1 FMI 12108 or $ new, Innova Biplane M4 BIP_4.18.5, DL_4.15.1 FMI 12108
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.