GE Healthcare · MRExciteSigna224916326489
Low Voltage on the High Voltage SSM Supply.
Symptom
Getting SSM Error When Auto Prescanning
Error message
2249163 SSM fault. High voltage power supply low. 2249163 Host: mr01 Process: NSP File: SCP:../RfNbMdsHWO.cpp Line: 2370 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: Gradient Echo Coil: 8HRBRAIN Scan: 26489/1 SSM fault.^A High voltage power supply low. 2249163 Host: mr01 Process: NSP File: SCP:../RfNbMdsHWO.cpp Line: 2370 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: Gradient Echo Coil: 8HRBRAIN Scan: 26489/1 SSM fault. High voltage power supply low.
Applies to
- Models
- Excite, Signa
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- SecondOpinion System Support Module 2160200-43 See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.