GE Healthcare · CTBrightSpeedVCT26010720426010720650180
BrightSpeed Select or Delight Select
Symptom
Gantry does not rotate. Customer noticed that gantry does not rotate.
Error message
Error: 260107204 Axial Motor Control (AMC) logic could not configure the AXIAL drive via the CAN network during configuration. Reset scan hardware to retry. See Long Description for full error Wed Apr 7 21:18:13 2004 Error: 260107204 Host: tgp Process: tgp_subsys File: amc_mgr.c Line: 230 Wed Apr 7 21:18:13 2004 Host : Table Gantry Processor Ermes # : 260107204 Exception Class : Abort Severity : Secondary File : amc_mgr.c Line# : 230 Function : AXIAL CONTROL : Axial Communications Scan Type : None/Unknown Scan: 0/0/0 Axial Motor Control (AMC) logic could not configure the AXIAL drive via the CAN network during configuration. Reset scan hardware to retry. Error: 260107206 Axial Motor Control (AMC) logic will NOT accept commands until the Axial Drive is properly configured/initialized. See Long Description for full error Wed Apr 7 21:04:02 2004 Error: 260107206 Host: tgp Process: tgp_subsys File: amc_mgr.c Line: 3771 Wed Apr 7 21:03:47 2004 Host : Table Gantry Processor Ermes # : 260107206 Exception Class : Abort Severity : Primary File : amc_mgr.c Line# : 3771 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Axial Motor Control (AMC) logic will NOT accept
Applies to
- Models
- BrightSpeed, VCT
- Product
- BrightSpeed Select, Delight Select or MERI Select Elite-16, Edge-8, Excel-4 HP_S_G16_G_HLT HP_S_G8_G_HLT HP_S_G4_G_HLT BrightSpeed Select or Delight Select (Hivalue Lite) 4/8/16 Slice Scanner BrightSpeed Delight Select Systems started to ship in first Quarter or 2009 with AIO (All in One) Console. AIO console uses HP8400 and NO DARC or IG's. Recon time is 6 FPS. If customer orders Fluoro option system requires one IG. The 16 FPS Upgrade in Not Available for Select systems. BrightSpeed Meri Select Systems ( Pancake Das ) started to ship in 3rd quarter of 2010, 4/8/16 Slice Scanner HiValue 16 Lite −−> BrightSpeed Elite Select HiValue 8 Lite−−> BrightSpeed Edge Select HiValue 4 Lite−−> BrightSpeed Excel Select BrightSpeed Select Scope: - 3.5MHu Solarix tube - JEDI 60 dc Lite - Modified NP Table Key Features: - 16x0.625mm(16x1.25) for Lite 16 - 8x1.25/8x2.5mm for Lite 4/8 - Rotation Speed: 0.8 s - Gantry: 2040mm width, 70cm Bore - Table: 1337@ISO-140mm scannable range - Recon: 3fps (w/o IG @ 2G Memory(4/8), 4G Memory(16)), optional: 6fps(w/ 1IG ) - Power: 42KW JEDI power See the attached Service Intro Planner for details Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-170021P15 (250V,8A) Fuse used in axial dynamic brake circuit as CT2A1A10F1 and found in FAN ASSY- GANTRY, and F4, F5 on HEMRC Interface Board and other places See long description for details Part #: 46-170021P15 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: 008.000A 0250V 3AB SLOW UL/ FRU: Yes Status: Orderable Comments: FLS - FUSE 8 AMP - 250 VOLT 3AB
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required