Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova16000331600032

Xray not Available - Reset System; If Problem Persists Call Service is displayed on the Image Monitor.

Symptom

Xray not Available - Reset System; If Problem Persists Call Service is displayed on the Image Monitor. Xray not Available - Reset System; If Problem Persists Call Service is displayed on the Monitor.

Error message

1600033 A possible hardware error has been detected by the IP disks chain AR XRImProc acquisition Component(s) DL Image Disk For Innova X100 the debug test is "MSA scsi disk 1 not ready : sense_data = 0x20400 - state = 0x20" 1600033 A possible hardware error has been detected by the IP disks chain AR XRImProc acquisition Component(s) DL Image Disk This error is due to errors found by the IPDisks driver. For Innova X100 the debug test is "MSA scsi disk 1 not ready : sense_data = 0x20400 - state = 0x20"

1600032 A software error has been detected by the Disk driver. The FAT loaded from the Disk is inconsistant 1600032 A software error has been detected by the Disk driver DO XRImProc acquisition Software_Problem 1 3 checkFat() failed => The FAT loaded from the Disk is inconsistant driver.c 4064

Applies to

Models
Innova
Product
Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova 2000 Innova 2000 systems Innova 2000 systems Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
Component
DL IMAGE DISKS DL IMAGE DISKS see long DL 36GB IMAGE DISKS 5144906 SCSI drive DL1 Image Disks Kit includes (2) 36GB image drives (ST336705LW), rack, fans, and cage. If no stock check part 2386702, also check SN4568B for alternate parts. P.Liebau 05/2009 DL 72GB IMAGE DISKS 5129897 *** OLC 2nd Opinion Part *** Kit contains (2) 72GB drives along with the mounting bracket, SCSI cable and terminator, for DL5+ and DL6 IP image disks. 1 FRU replaces both image disks. Contact your local OLE for approval. Replaced by new part 5248652 but that does not include the SCSI cable. P.Liebau 02/2009 5199031 See SN4844 for kit contents. Replaces 2332254 and 2386700 for DL2 to DL6. P.Liebau 11/2010 5248652 Kit contains (2) 72GB image disks and SCSI terminator, does NOT include SCSI cable which is part of DL internal cable set 5129896. P.Liebau 02/2009
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare X-ray Vascular issues