GE Healthcare · MRExcite2260007
Scans Are Slow To Start due to Nonexistent High Order Shim Supply (HOS).
Symptom
Scans Are Slow To Start System hesitates before starting to scan on all protocols.
Error message
Resetting shim supply, please wait See Long Description "Resetting shim supply, please wait" This message appears to the operator in the message box area. This message is not logged. 2260007 Time out occurred while accessing the shim supply
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.