Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova3602045600105700105100002536020513702051

Collision/patient fall risk.

Symptom

Collision/patient fall risk. Alarm system failureArcnet communication error issue. error Code = 600105, 1000025, Error message; Collision/patient fall risk. Alarm system failure; Arcnet communication error issue. error Code = 600105, error Text = ARCNET Communication Failure followed by error Code = 1000025, error Text = ARCNET Communication Failure. Mainly due to Positionner, can be seen in application or in boot phasis. Positionner is not there, and both RTAC SW actors managing POS (XRImChainPOS) and collimator (XRbeamform) try to communicate to POS every minutes. Error message; Collision/patient fall risk. Alarm system failure; Arcnet communication error issue. error Code = 600105, error Text = ARCNET Communication Failure followed by error Code = 1000025, error Text = ARCNET Communication Failure. Mainly due to Positionner, can be seen in application or in boot phasis. Positionner is not there, and both RTAC SW actors managing POS (XRImChainPOS) and collimator (XRbeamform) try to communicate to POS every minutes.

Error message

3602045, error Text = Jedi: Inverter Gate Power Supply error, debug Text += JEDI Error Code: 0503H Inverter Gate Power Supply error, recovery Class = IH, major Function /= XRTechMgr System impact: X-ray unavailable. Displayed message : "Xray not available - Reset system; if problem persists, call Service".

error Code = 600105 (FRT) and/or 700105 (LAT), error Text = ARCNET Communication Failure , debug Text /= Null , recovery Class = IH, major Function /= XRBeamForm Error Code = 600105 (FRT) and/or 700105 (LAT), error Text = ARCNET Communication Failure , debug Text /= Null , recovery Class = IH, major Function /= XRBeamForm System impact: X-ray and Gantry motion unavailable.

Error Code = 1000025, error Text = ARCNET Communication Failure, debug Text = Positioner link disconnected, recovery Class = IH, major Function /= XRImChainPos Error Code = 1000025, error Text = ARCNET Communication Failure, debug Text = Positioner link disconnected, recovery Class = IH, major Function /= XRImChainPos. System impact: X-ray and Gantry motion unavailable. Displayed message : "Xray not available - Reset system; if problem persists, call Service".

error Code = 3602051 front, 3702051 Lat: Jedi: C

Applies to

Models
Innova
Product
Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare X-ray Vascular issues