GE Healthcare · MRExciteDiscoveryOptima222385010303224935200000046000000102249425
CWA ID135168. Auto and Manual Prescan Fails.
Symptom
Auto and Manual Prescan Fails
Error message
2223850 Content Of Packet Cant Be Routed To A Host Application Thu Feb 2 11:01:58 2006 Error: 2223850 Host: RIMIMR Process: NSP File: PMH.cc Line: 456 Contents of packet can't be routed to a host application: opcode:10303 seq:20 rev:2 type:1 length:8 stat:-1 resp:WavRec req:HC 2249352 Object: APS. Maximum CM connect failure with these arguments: (00000046, ee 02 2707 0a000202, 00000010) 2249425 A network error was detected by task tCM_TxD1 2249425 Object: APS.A network error was detected by task tCM_TxD1 (0x673A090). Additional information in /usr/g/service/log/vxNetLg-27af7-> 2228179 TPS detected NSP (host) timeout Tue Sep 20 14:17:14 2005 Error: 2228179 Host: lx-mr Process: NSP File: APS:hs watchdog check Line: 10 TPS detected NSP (host) timeout and terminated any scan in progress.
Applies to
- Models
- Excite, Discovery, Optima
- Product
- 3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
- Software
- 14.0 M5 Excite HDx HDi Platforms only.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.