GemnetIV, getting DVD library hardware fault.
Symptom
GemnetIV, getting DVD library hardware fault. GemnetIV is showing error DVD library hardware fault. Several patients were marked offline. If you selected these patinets that were on a DVD disk, they had a red circle around the DVD icon in GEMNETIV Explorer. So the library could not pull any of the patients. One thing that I found was that the diskextender Agent was not running. The diskextender Server was. What made me look at this was the diskender was showing the extended disk drive offline. After restarting this service, I ran dbaddotg - Several patients were marked offline. If you selected these patinets that were on a DVD disk, they had a red circle around the DVD icon in GEMNETIV Explorer. So the library could not pull any of the patients. One thing that I found was that the diskextender Agent was not running. The diskextender Server was. What made me look at this was the diskender was showing the extended disk drive offline. After restarting this service, I ran dbaddotg -v again. After it ran, there were no errors. The red circles were gone and you could now get to these patinets. System is up and running.
Error message
Gemnet IV, Not retreiving patient information off of quantum Tape drives. Defective Tape Drive Not retreiving patient information off of quantum Tape drives. Defective tape drives Gemnet IV, Not retreiving patient information off of quantum Tape drives. Defective tape drives
Applies to
- Product
- Gemnet Image Vault Gemnet Image Vault Gemnet Image Vault
- Software
- Gemnet 2.1 D sp3 Gemnet 2.1 D sp3 Gemnet 2.1 D sp3
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.