GE Healthcare · MRExcite22521062219999200002368
CWA ID134909. Application Software does not complete the boot process with TPS Download failure.
Symptom
Application Software does not complete the boot process with TPS Download failure
Error message
2252106 Failed to determine hardware state for SCP board within 20 seconds 2252106 Failed to determine hardware state for AGP board within 20 seconds Error: 2252106 Host: rmns Process: NSP File: diags_utils.c Line: 666 Diagnostic software error. Failed to determine hardware state for AGP board within 20 seconds The board may be busy loading vxWorks or may be hung. 2219999 Internal Software Error in read protocol interpreter: /usr/g/protocols/, Tcl_EvalFile. 200002368 A set of processes are deadlocked waiting for each other to start. Mon Apr 12 11:43:42 2004 Error: 200002368< Host: BaptistGEMR Process: hostMonitorbsp File: ssProcStart.c Line: 1099 A set of processes are deadlocked waiting for each other to start. svat, scn, rtcProxy, resetServer, lais, hgp, displayServer, autoview, app, WLServer, TIR, SITT, RtpHost, RTPSIServer, RIR, PPSproxy are the likely culprits. This deadlock has caused the system to shutdown!
Applies to
- Models
- Excite
- Product
- Open Speed with EXCITE For HFO4 and HFO5 only
- Software
- OPENMR4-HFO4 Software version for HFO4 system, which is OpenSpeed with Excite Technology
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.