GE Healthcare · MRExcite
Coil ID LED is off when 3T NV array coil is connected due to defective Coil cable.
Symptom
Coil ID LED does not illuminate when 3T NV Array coil is connected Coil ID LED is off and the system does not acknowledge that any coil is connected.
Applies to
- Models
- Excite
- Product
- 3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
- Software
- 14.0 M5 Excite HDx HDi Platforms only.
- Component
- 2414385 E3 SYSTEM CABLE, 3T NV
- Area
- MR