The helium level is dropping greater than 5% over a 30 day period due to a possible helium leak.MAC00005 Proactive Magne…

The helium level is dropping greater than 5% over a 30 day period due to a possible helium leak.MAC00005 Proactive Magnet Performance Issue.Low level Helium Loss > 5% in 30 days.

Symptom

MAC00006 Proactive Magnet Performance Issue. Low level Helium Loss > 5% in 30 days. The OnWatch engine has determined that the helium level has experienced a low level helium loss greater than 5% in the past 30 days. NOTE: The Digital Services (DS) Rules Engine detected this problem. This issue is only applicable when detected in this way, and should not be reused in Knowledge objects for normal reactive service.

Error message

MAC00006 Proactive Magnet Performance Issue. Low level Helium Loss > 5% in 30 days. The OnWatch engine has determined that the helium level has experienced a low level helium loss greater than 5% in the past 30 days. NOTE: The Digital Services (DS) Rules Engine detected this problem. This issue is only applicable when detected in this way, and should not be reused in Knowledge objects for normal reactive service.

Applies to

Product
Magnet LCC. Software Version Concept should NOT be selected for Magnet Products Serial #: 1.5T (Rxxx) / 1.0T (Qxxx) He Vessel Pressure:>3 psig but <5 psig Shield Temp:< 55 K Coldhead Temp:> 3.8K but < 4.5K Recondenser Temp:> 4.2K but < 4.55K Delta Temp (Coldhead - Recondensor):> 0 but < 0.3K Min. LHe Fill Level: 50% Min. LHe Ramp Level: 90% Min. LHe Shim Level: 75% Helium Volume:1940 L (1.5T) Helium Volume:1970 L (1.0T) Burst Disk <":2319285 Burst Disk >":2350415 Burst Disk O-ring:46-281101P1 Baffle Assy 2180818 Low Cryogen Consumption
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required