CWA ID133982. Some or All Images in the Last Scan Did Not Reconstruct.
Symptom
Some or All Images in the Last Scan Did Not Reconstruct Some or All Images in the Last Scan Did Not Reconstruct.
Error message
2218077 Cannot find shadow path for image #2 Wed Oct 27 08:28:18 2004 Error: 2218077 Host: lx-mr Process: ifcc File: ImagePath.c Line: 43 Cannot find shadow path for image #2
Applies to
- Models
- Signa, Ovation, Excite
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution