Unable to scan with Power Monitor Faults on Head Coil due to Center Frequency Was Off By 400 Hz.
Symptom
Unable to scan with Power Monitor Faults
Error message
2254782 The Universal Power Monitor 1 has detected Peak Power Exceeded on the Narrow Band Head Amplifier. 2254782 NSP SCP:UPM 657 The Universal Power Monitor 2 has detected Peak Power Exceeded on the Narrow Band Head Amplifier. Peak value detected = 429 W. Max Peak Limit = 426 W- Possible cause(s): ASC RF Detector, ASC RF Processor, RF Amp, Software .
Applies to
- Models
- Excite, Signa
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution