GE Healthcare · MRExciteDiscoveryOptima222381722521062223851224715611409200002352
CWA ID133135. Application Software does not complete the boot process with TPS Download failure Due to Idle Overnight Wi…
CWA ID133135. Application Software does not complete the boot process with TPS Download failure Due to Idle Overnight With Power On.
Symptom
Application Software does not complete the boot process with TPS Download failure
Error message
2223817 TPS is Not Responding. Please Reset TPS Wed Apr 28 16:58:01 2004 Error: 2223817 Host: GEMS Process: NSP File: TardisReset.cc Line: 398 TPS is not responding. Please reset TPS. 2252106 Failed to determine hardware state for SCP board within 20 seconds 2252106 Failed to determine hardware state for AGP board within 20 seconds Error: 2252106 Host: rmns Process: NSP File: diags_utils.c Line: 666 Diagnostic software error. Failed to determine hardware state for AGP board within 20 seconds The board may be busy loading vxWorks or may be hung. 2223851 Reset TPS was not Successful. errno 11 -> 2223851 1 MROC_OC0 NSP TardisReset.cc 471 Reset TPS was not successful. errno : 11-> Note: The extended log below can be viewed from the Service Browser with the Service Key installed. Pull up the GE System log under the Error Logs tab. If there is an extended log available, the error number will be a hyperlink. Clicking on it will reveal the extended log below. 2223851 ERROR: EM_mrNSP_ITLink_RESET_FAILED Reset TPS was not successful. errno : %d Extended Error: The MGD failed to start a TPS Reset. Either the serial connection to the MGD STIF Board is not working or the Ethernet connection
Applies to
- Models
- Excite, Discovery, Optima
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.