GE Healthcare · X-ray VascularInnova1050305
The Omega table will not drive (move) up or down. all other table and gantry motions work.
Symptom
The Omega table will not drive (move) up or down. all other table and gantry motions work. The Table does not drive up or down when controlling the elevation from the TSUI (smart Box/handle). The Table does not drive up or down properly, when controlling the elevation from the TSUI (smart Box/handle) or the elevation drive is intermittent.
Error message
error Code = 1050305, error Text = TUBE COLL NOT CLEARED, debug Text = - TUBE COLLISION NOT CLEARED. CONTINUE WITH CAUTION (T:4A14). In MP_SetCollisionNotClearedMsgOnIRM, recovery Class = WR, major Function /= GantryAxis System impact: Half speed until system collimator level1 collision is cleared. Displayed message : "Tube collision not cleared. Please clear collision or call Service.". Root Cause: TUBE COLLISION NOT CLEARED This error is due to clear timeout after a collision detected on the Tube side. Recommended Action: The expected behaviour of the collision detection is as follows: When a level 1 or level 1bis bump is detected, then the positioner software stops all motions until the operator releases the enable. When the operator presses again on the gantry enable, then the gantry shall move as follows: -if a collision is detected on the tube, then the gantry plays back ("backout motion"), for a maximum of 2 seconds, its last trajectory before the collision detection until the collisions are cleared. If trajectory is not sufficient and collisions are still active at the end of the backout motion (abnormal end), then a “COLLISION NOT CLEARED” message is displayed and the manuApplies to
- Models
- Innova
- Product
- Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova 2000 Innova 2000 systems Innova 2000 systems Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
- Component
- Positioner Gantry Positioner Gantry Positioner Gantry
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.