Pular para o conteúdo principal
GE Healthcare · X-ray VascularAdvantxInnova

Smart Handle replacement

Symptom

C-arc not moving Customer said the gantry motion did not work anymore. Then, he told that initial problem was located on the C-arc motion. He was unable to drive C-arc (motion starts then automatically stops). If L-arm at 0°, only C-arc motion is impacted whatever machine / anatomical mode selected on Smart handle / smart Box. if L-arm is moved to a position different from 0°, C-arc AND pivot motion will be impacted in ANATOMICAL mode. Customer said the gantry motion did not work anymore. Then, he told that initial problem was located on the C-arc motion. He was unable to drive C-arc (motion starts then automatically stops). If L-arm at 0°, only C-arc motion is impacted whatever machine / anatomical mode selected on Smart handle / smart Box. if L-arm is moved to a position different from 0°, C-arc AND Pivot motion will be impacted in ANATOMICAL mode.

Error message

4B07 with hardware 104 and user 0002 indicates rocker c arc control is defective. Attempt to calibrate rocker. If this fails replace smart handle. 4B07 with hardware 104 and user 0002 indicates rocker c arc control is defective. Attempt to calibrate rocker. If this fails replace smart handle.

Applies to

Models
Advantx, Innova
Product
Advantx E Monoplane Advantx E Systems advantx e, vamp Advantx E generator Systems Advantx E Biplane Advantx E biplane systems: LCLP+ (22 cm II), LCN+ (32 cm II), LCALP+ (40 cm frontal plane, 32 cm lateral plane) Advantx E biplane systems: LCLP+ (22 cm II), LCN+ (32 cm II), LCALP+ (40 cm frontal plane, 32 cm lateral plane) Advantx, LFX Advantx product line LFX Advantx product line
Component
smart handle smart handle smart handle one type of handle is: 5145783-2 FRU Smart handle Zenith_Unity 212000947 09/22/2010 10:28 GMT see vascular suport central site (@vascparts) for a cross refference of handles, boxes, and tssc's 212000947 04/09/2010 18:48 GMT Positioner Gantry Positioner Gantry Positioner Gantry
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required