GE Healthcare · X-ray VascularInnova70082
132242 Error CANNOT SWITCH RECEPTORS DURING AN ACQUISITION on monitors
Symptom
Error CANNOT SWITCH RECEPTORS DURING AN ACQUISITION on monitors
Error message
70082 CANNOT SWITCH RECEPTORS DURING AN ACQUISITION HWStatus=0000 User=0000 0000 6700 70082 : Recommended Actions: Root Cause : Software problem. This error is due to parameter out of range, bad state, bad pointer detected in the Titan application software. On the Advantx an attempt was made to switch receptors at the console or while fluoro or record exposures are in progress. Recommended Actions: If an attempt was made to switch receptors at the console or while fluoro or record exposures are in progress wait until exposures are complete before attempting the switch again. If recovery class of error is IH or AR , then reload Titan subsystem software. If problem persists, then replace Titan. If recovery class is DO, then it could be only an effect of other problem.
Applies to
- Models
- Innova
- Product
- Innova 2000 Innova 2000 systems Innova 2000 systems
- Software
- $, Innova 2000 Service Pack FMI F4, HerSvcPack_2.02 HerSvcPack_2.02 DL Applications version 11.23.26 Ghost DL-1 "F" : 17.4 Ghost DL-1 "G" : 15.5 Ghost DL-2 : 16.4 Ghost DL-4 : NA DL : 11.23.26 DL Atlas Core : 3.1O-1 RTAC Atlas Core : 3.1O-2 IDC : 2.13 Calypso : 11.13_FMI_7 Calypso Uptime : 11.25_FMI Titan : H0.57 Pos. Motorola : 5.2Y3-2 Pos. Roadrunner : 5.0R
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.