Swapped labeling of J11 and J12 connectors of fiber optic cable run causing the installer to cross the connections going…
Swapped labeling of J11 and J12 connectors of fiber optic cable run causing the installer to cross the connections going from the pen-panel to the SRI.
Symptom
SRI3 Loop back Test Fails Effectivity: New EXCITE III installs. Only 30 Magnets shipped are possibly affected. Problem Description: During installation, unable to communicate with SRI3 - SRI loop back test fails. SMILe 2/4/05
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution